ACCURACY
3-2
IMPEDANCE ACCURACY
Accuracy of Z = ± [A x K
i
x K
v
+ (K
h
+ K
l
) x 100] %
A = Basic Impedance Accuracy Factor
from graph on next page.
If the unit is in the constant voltage
(CV) mode, double the basic
accuracy factor.
K
i
= Integration Time Factor (see K
i
Table below)
K
v
= Drive Voltage Error Factor
(see K
v
Table below)
K
h
, K
l
= Extreme Range Error Term (see
K
h
,K
l
Table below)
K
i
Table
Meas Rate Frequency
Z
m
K
i
slow,med all 1
fast 100 Hz-1 kHz 6.25 <Z
m
<400 k 3
fast all other 2
K
v
Table
V
out
(V rms) K
v
1.0 - 0.55 1 / V
out
0.5 - 0.3 0.5 / V
out
0.25 - 0.15 0.25 / V
out
0.10 0.11/V
out
Note that K
v
equals one for the primary drive
voltages (1.0, 0.5 and 0.25 Vrms).
K
h
& K
l
Table
Frequency K
l
K
h
100,120,1k Hz (1 m / Z
m
)(Z
m
/ 2 G)
10 kHz (1 m / Z
m
)(Z
m
/ 1.5 G)
100 kHz (4 m / Z
m
)(Z
m
/ 50 M)
Z
m
= the impedance of the device being
measured.
Note that K
l
is negligible for impedances above
100 and K
h
is negligible for impedances below
1 k, both at all frequencies.